This product is specially used for the atomization treatment of respiratory tract, oral cavity and lung, and the portable humidification and moisturizing of baby's sleep. It adopts the latest technology and has three charging display lamps, which will be always on when full
The specific parameters are as follows:
1. Battery type: rechargeable polymer lithium battery/2 x AA Cell/Type-C plug in power supply
2. Fuselage size: length, width and height:43.8mmx43.8mm x120mm
3. Fuselage weight: 83g
4. Package size: 124mmx146.3mmx51.8mm
5. Particle Size: 4um
6. Product Name:Nebulizer
7. It Includes:Nebulizer,Mask,Mouthpiece,USB Cable,Manual
In normal atomization
The operation key LED1 is always on
Water shortage in atomization
The atomization is automatically stopped, and the main light flashes slowly for 3 times, then it will automatically shut down and enter sleep
Remove the atomizing head during atomization
Atomization low battery
In normal atomization
The operation key LED1 is always on
Water shortage in atomization
The operation key LED1 is always on
The atomization is automatically stopped, and the main light flashes slowly for 3 times, then it will automatically shut down and enter sleep
Remove the atomizing head during atomization
Maintenance
This product belongs to the micro mesh atomizer. The biggest difference between
this product and the air compression atomizer lies in the noise, but it also has its shortcomings. The biggest problem is the blocking of the network. When using this kind of product, please be sure to maintain it according to the following use methods
After receiving the goods, dismantle the atomizing head in time, wipe the atomizing head with a wet towel, put water into it, install it on the host, and atomize it for three minutes. The purpose is to prevent fine dust in the production and packaging process and keep the micropores of the atomizing head unobstructed and clean;
When the atomizing head container is clean and dry (without water droplets), it can be atomized after putting the liquid to be atomized;
After using the special liquid atomization, put in clean water, atomize for three minutes, pour out the excess clean water, wipe the atomizer container dry, keep it dry, cover it and store it for next use;
matters needing attention:
Do not retain special liquids, such as normal saline and special medicine liquid, in the atomizer container for a long time to form small particles and cause the atomization hole to be blocked, resulting in small or no fog;
It is not recommended to leave the main engine idle after removing the atomizing head, which will greatly shorten the service life of the main engine;